High-Purity Germanium Gamma Spectrometer

The HA4200-HPGe is a laboratory-grade gamma spectrometer integrating a high-purity germanium crystal, advanced preamplifier, low-background lead shield, high-stability Digital Multichannel Analyzer (MCA), and cryogenic cooling system. Paired with fully-featured gamma spectrum analysis software, it enables precise measurement and characterization of radioactive samples.
Add to inquiry

Product Overview

The HA4200-HPGe is a laboratory-grade gamma spectrometer integrating a high-purity germanium crystal, advanced preamplifier, low-background lead shield, high-stability Digital Multichannel Analyzer (MCA), and cryogenic cooling system. Paired with fully-featured gamma spectrum analysis software, it enables precise measurement and characterization of radioactive samples.

 

Applications

This instrument is used for radioactivity measurement of nuclear power, low-level samples (environmental, biological, food), inspection and quarantine, biomedicine, astrophysics and chemistry, geology, forensic science, archaeology, metallurgy, materials science, and many other scientific and social fields. The high-purity germanium gamma spectrometer is also specifically designed for nuclide identification and radiochemical purity testing of radiopharmaceuticals.

 

Features

  • Proprietary Spectrum Analysis Software:​ Integrates acquisition, control, analysis, and automated reporting. User-friendly interface supports one-click report generation.
  • Flexible Cooling System:​ Optional liquid nitrogen (LN₂) dewar or closed-cycle electric cooler. The recondensing system extends LN₂ refill intervals up to 24 months.
  • Low-Background Lead Shield:​ Background count rate < 2 cps (50 keV–2 MeV @ 50% relative efficiency). Features a sliding-top access door for easy sample placement.
  • Configurable Digital MCA:​ Optional 16k, 32k, or 64k channels to meet specific resolution requirements.
  • Source-Free Efficiency Calibration:​ Performs efficiency calibration for point, volumetric, planar, and Marinelli beaker geometries without standard sources. Supports elemental, compound, and mixed matrices; calculation error ≤ 3%.
  • Low Acoustic Noise:​ Noise level < 60 dB(A) at 1 m during full-power operation.
  • 3D Laser Scanning Modeling:​ Creates accurate 3D geometry models for import into the efficiency calibration software, ensuring high-accuracy results.

 

System Configuration

  • HPGe Detector
  • Low-Background Lead Shield
  • Digital Multichannel Analyzer (MCA)
  • Gamma Spectrum Analysis Software
  • Source-Free Efficiency Calibration Software
  • Cooling System
  • 3D Laser Scanner (Optional)

 

Technical Specifications

Product Model

RPPX-30185 (Broad-Energy Type)

Energy Range

5 keV ~ 10 MeV

Relative Efficiency

≥ 30% (Higher efficiency options available)

Window Material

Beryllium or Carbon Fiber

Energy Resolution

≤1000 eV @ 122 keV (⁵⁷Co)

≤1.85 keV @ 1.33 MeV (⁶⁰Co)

Peak-to-Compton Ratio

58:1

Peak Shape Parameters

FW0.1M/FWHM = 1.9

FW0.02M/FWHM = 2.65