The HA4300-R is an intelligent, precision dose calibrator specifically designed for nuclear medicine, enabling rapid quantification of radiopharmaceutical activity. It features one-touch background subtraction and a user-friendly full Chinese interface for exceptional ease of use.
Add to inquiry

Product Overview

The HA4700 Thyroid & Lung Organ Counter comprises thyroid detectors, lung detectors, gamma spectrum analysis software, internal dose assessment software, optional source-free efficiency calibration software, a computer, and an integrated shielded measuring chair. It simultaneously quantifies radionuclide activity in the human thyroid and lungs, calculating internal exposure doses based on the measurement results.

 

Applications

Designed for on-site organ radioactivity screening in nuclear emergency response, nuclear power plants, spent fuel reprocessing facilities, and special nuclear material processing sites. Also deployed as emergency equipment for Centers for Disease Control and Prevention (CDC).

 

Features

  • Native Chinese Software:​ Full Chinese interface for spectrum analysis and internal dose evaluation.
  • Advanced Spectral Analysis:​ Automatic spectrum stabilization, energy calibration, peak searching, multiplet analysis, activity calculation, and measurement uncertainty analysis.
  • Flexible Control:​ Adjustable gain and high voltage; real-time display of live time, real time, and discriminator threshold.
  • Extensible Nuclide Library:​ Preloaded standard radionuclide library with support for user-defined nuclides.
  • Source-Free Efficiency Calibration:​ Built-in algorithm for self-calibration of gamma detection efficiency without standard sources.
  • Comprehensive Dose Calculation:​ Estimates internal organ doses considering nuclide type, metabolic half-life, age, and specific organ/tissue parameters.
  • Automated Reporting:​ Auto-generates test reports with support for external printing.
  • Ergonomic Design:​ Chair-type measurement station with adjustable height and detection angle for subject comfort.

 

Technical Specifications

Product Model

HA4700

Detector Type

NaI(Tl) Scintillation Crystal

NaI(Tl) Scintillation Crystal

Detector Size

Φ1.5 in × 1.5 in

Φ5 in × 4 in

Energy Range

50 keV ~ 3 MeV

50 keV ~ 3 MeV

Energy Resolution

≤7.5%

≤8%

MCA Channels

1024

1024

Minimum Detectable Activity

(Specified counting time)

¹³¹I ≤ 125 Bq, ¹³³I ≤ 180 Bq (60 s)

¹³¹I ≤ 80 Bq, ¹³³I ≤ 100 Bq (180 s)

¹³⁷Cs ≤ 95 Bq (180 s)

 ⁶⁰Co ≤ 136 Bq (180 s)

Measurement Speed

3 min per subject

3 min per subject

Power Supply

AC 220 V, 50 Hz

AC 220 V, 50 Hz

Operating Temperature

−10 °C ~ 50 °C

−10 °C ~ 50 °C

Operating Humidity

≤ 90% RH