Home    Radiation Monitoring Network    Radioactive Aerosol & Iodine Monitor

Radioactive Aerosol & Iodine Monitor

Add to inquiry

Product Overview

The HA1600 Radioactive Aerosol & Iodine Monitor uses a vacuum pump to extract air from the target area. Aerosols and iodine in the air are collected on filter paper and trapped in an iodine cartridge. The activity of the collected filter paper and iodine cartridge is then measured to analyze the radioactivity levels of aerosol particles and gaseous iodine in the air.

 

Features

  • Multiple Measurement Modes: Simultaneously measures the activity levels of aerosols and iodine in the air.
  • Covers Common Nuclides: Covers common medical gamma-emitting nuclides; expandable nuclide library as required.
  • Excellent Shielding Performance: Equipped with a lead chamber with shielding efficiency >95% to minimize environmental interference.
  • Complete Monitoring Records: Measures in compliance with standard procedures, stores detailed spectra and historical records, meeting regulatory requirements.
  • High Detection Accuracy: Employs spectrum stripping, sample deposition correction, decay correction, and radon and thoron background correction to ensure measurement accuracy.
  • Safety Emergency System: Monitors pump operation in real time; stops the pump immediately in case of sampling abnormality to improve equipment safety.
  • Integrated & Movable: Integrated and movable structure; sampling gas line compatible with various applications, supporting multi-point mobile monitoring for different users.

 

Technical Specifications

Product Model

HA1600

Detector Type

NaI(Tl) crystal

Detector Size

100 nSv/h ~ 100 mSv/h

Relative Intrinsic Error

2 inches

Detectable Nuclides

⁶⁷Ga, ¹²⁵I, ¹³¹I, ⁹ᵐTc, ²⁰¹Tl, single-photon emitters (editable & expandable)

Lead Chamber Shielding Efficiency

>95%

Sampling Flow Rate

0–50 LPM

Measurement Range

3.7 ~ 3.7×10⁶ Bq/m³

Energy Resolution

≤ 7.5% (@ 662 keV, ¹³⁷Cs)

MCA Channels

2048

Power Consumption

≤ 250 W